FDA Adverse Event Malfunction Summary report: N

HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE

MDR report key: 3232458 · Received July 18, 2013

Report

Report Number
2024168-2013-04495
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 24, 2013
Report Date
June 24, 2013
Manufacturer
AV-TEMECULA-CT
Product Code
DQX
PMA / PMN Number
K103101
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR ANALYSIS. THE SEPARATION WAS ABLE TO BE CONFIRMED. BASED ON A VISUAL INSPECTION AND SCANNING ELECTRON MICROSCOPY (SEM) IMAGING OF THE RETURNED DEVICE, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE REVIEWED INFORMATION, NO PRODUCT DEFICIENCY WAS IDENTIFIED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT AN UN-SPECIFIED LESION IN THE PERIPHERAL. THE BALANCE MIDDLEWEIGHT (BMW) ELITE GUIDE WIRE WAS ADVANCED WITHOUT ISSUE. A NON-ABBOTT EMBOLIC PROTECTION DEVICE WAS THEN ADVANCED, AND THE BMW ELITE GUIDE WIRE WAS REMOVED. AFTER REMOVAL, IT WAS NOTED THAT A PORTION OF THE GUIDE WIRE HAD SEPARATED. THERE WAS NO RESISTANCE NOTED DURING REMOVAL OF THE GUIDE WIRE. THE SEPARATED PORTION WAS REMOVED WITH THE EMBOLIC PROTECTION DEVICE. A NEW GUIDE WIRE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE PATIENT EFFECTS AND NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
333150 HI-TORQUE BALANCE MIDDLEWEIGHT ELITE GUIDE WIRE GUIDE WIRE DQX AV-TEMECULA-CT 2081001

Patients

Seq Age Sex Outcome Treatment
1 73 YR EMBOLIC PROTECTION: SPIDER