FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3232454
·
Received July 18, 2013
Report
- Report Number
- 1416980-2013-18785
- Event Type
- Malfunction
- Date Received
- July 18, 2013
- Date of Event
- June 10, 2013
- Report Date
- June 10, 2013
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE AND THE SAMPLE WAS VISUALLY INSPECTED. THE REPORTED CONDITION OF THE PLUG ICON NOT ILLUMINATING WAS CONFIRMED. THE CAUSE OF THE REPORTED CONDITION WAS DUE A DAMAGED DC POWER SUPPLY. TO CORRECT THE ISSUE THE POWER SUPPLY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP PLUG ICON DID NOT ILLUMINATE WHEN IT WAS PLUGGED IN. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 334694 | FLOGARD | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |