FDA Adverse Event Malfunction Summary report: N

MIS SINGLE INNER SET SCREW

MDR report key: 3232417 · Received July 18, 2013

Report

Report Number
1526439-2013-21467
Event Type
Malfunction
Date Received
July 18, 2013
Date of Event
June 25, 2013
Report Date
June 25, 2013
Manufacturer
DEPUY SYNTHES SPINE
Product Code
NKB
PMA / PMN Number
PK041801
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

A FOLLOW UP REPORT WILL BE FILED UPON COMPLETION OF THE INVESTIGATION.

Additional Manufacturer Narrative · 1

VISUAL EXAMINATION OF THE RETURNED SET SCREW CONFIRMED THE THREADS HAD TORN AWAY FROM THE DEVICE. A REVIEW OF THE DHR ACKNOWLEDGED THAT NO ISSUES WERE IDENTIFIED DURING THE MANUFACTURING AND RELEASE OF THIS PRODUCT THAT COULD HAVE BEEN ATTRIBUTED TO THE PROBLEM REPORTED BY THE CUSTOMER. THE PRODUCT WAS RELEASED ACCOMPLISHING ALL QUALITY REQUIREMENTS. NO DEFINITIVE CONCLUSIONS CAN BE DRAWN AS TO THE CAUSE THE TORN THREADS. HOWEVER, THE DAMAGE SUGGESTS THAT THE SET SCREW BECAME CROSS THREADED DURING ADVANCEMENT WITHIN THE CONCOMITANT DEVICE PEDICLE SCREW. IN THE ABSENCE OF AN IDENTIFIED DEVICE MANUFACTURING/RELEASE ISSUE OR OBSERVED TREND, THIS COMPLAINT WILL BE CLOSED WITH NO FURTHER ACTION REQUIRED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN MIS PROCEDURE, THE THREADS OF TWO SET SCREWS BECAME DEFORMED DURING ADVANCEMENT. AS A RESULT THE SET SCREWS COULD NOT BE ASSEMBLED TO MATING PEDICLE SCREW. ALSO, THE INNER THREADS OF THE MATING PEDICLE SCREW WERE ALSO DEFORMED. HOWEVER, RECEIPT OF THE COMPLAINT SAMPLES ON (B)(4) 2013 FOUND THREADS HAD BEEN TORN FROM THE SET SCREWS AND NOT SIMPLY DEFORMED AS REPORTED. CONCOMITANT DEVICE: PEDICLE SCREW, CATALOG NUMBER UNKNOWN. SEE MFG MEDWATCH REPORT NO. 1526439-2013-21468 FOR THE SECOND SET SCREW THAT WAS INVOLVED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335703 MIS SINGLE INNER SET SCREW ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE NKB DEPUY SYNTHES SPINE ANGBT6

Patients

Seq Age Sex Outcome Treatment
1