FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3232394 · Received July 18, 2013

Report

Report Number
2531779-2013-11114
Event Type
Malfunction
Date Received
July 18, 2013
Report Date
June 27, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANIMAS HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF ANIMAS RECEIVES THE PRODUCT ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/14/2013 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. DURING TESTING, THE PUMP WAS POWERED ON AND THE DISPLAY SCREEN APPEARED DISCOLORED. WHEN REPLACED WITH A TEST DISPLAY, THE SCREEN FUNCTIONED PROPERLY.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER ALLEGED THAT THE ISSUE OCCURRED GRADUALLY AND NOTICED THE ISSUE SINCE MARCH. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
335495 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 10 YR