FDA Adverse Event Malfunction Summary report: N

QUANTA SYSTEM S.P.A.

MDR report key: 3232324 · Received December 21, 2012

Report

Report Number
3004378299-2012-00015
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
June 21, 2012
Report Date
December 19, 2012
Manufacturer
QUANTA SYSTEM, S.P.A.
Product Code
GEX
PMA / PMN Number
K090962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THERE WAS A LOOSE POWER WIRE IN THE CHILLER 1. THE WIRE WAS RECONNECTED AND THE LASER SYSTEM RETURNED TO WORK. WE ARE UNAWARE ABOUT PT INJURY.

Description of Event or Problem · 1

THE CHILLER 1 SHOWED THE MESSAGE: FLOW SENSOR 1 ERROR. UNABLE TO MAINTAIN CHILLER TEMPERATURE SET POINT. WE ARE AWARE ABOUT PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTA SYSTEM S.P.A. SURGICAL LASER GEX QUANTA SYSTEM, S.P.A. CYBER TM 120

Patients

Seq Age Sex Outcome Treatment
1