FDA Adverse Event Malfunction Summary report: N

QUANTA SYSTEM S.P.A.

MDR report key: 3232316 · Received December 18, 2012

Report

Report Number
3004378299-2011-00018
Event Type
Malfunction
Date Received
December 18, 2012
Date of Event
September 21, 2011
Report Date
December 14, 2012
Manufacturer
QUANTA SYSTEM S.P.A.
Product Code
GEX
PMA / PMN Number
K100558
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE PROBLEM WAS IN THE FAN OF HEAT SINK SYSTEM, THE FAN WAS RESET AND THE LASER SYSTEM CORRECTLY WORKED. WE ARE UNAWARE ABOUT DELAY ON TREATMENT OR PT INJURY.

Description of Event or Problem · 1

THE LASER SYSTEM SHOWED TEMPERATURE ALARM AT 40 DEGREES CELSIUS. WE ARE UNAWARE ABOUT DELAY ON TREATMENT OR PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTA SYSTEM S.P.A. SURGICAL LASER GEX QUANTA SYSTEM S.P.A. IG-1470

Patients

Seq Age Sex Outcome Treatment