Report Number
3004378299-2012-00024
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
November 27, 2012
Report Date
December 20, 2012
Manufacturer
QUANTA SYSTEM, S.P.A.
Product Code
GEX
PMA / PMN Number
K090962
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE TWO CHILLER UNITS HAD USED UP THE THERMAL EXCHANGE GAS. THESE CHILLERS WERE REPLACED AND THE LASER SYSTEM RETURNED TO WORK. WE ARE UNAWARE ABOUT DELAY IN TREATMENT OR PT INJURY.

Description of Event or Problem · 1

THE TWO CHILLER UNITS DID NOT MAINTAIN THE TEMPERATURE SET POINT. WE ARE UNAWARE ABOUT DELAY ON TREATMENT OR PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTA SYSTEM S.P.A. SURGICAL LASER GEX QUANTA SYSTEM, S.P.A. CYBER TM 120

Patients

Seq Age Sex Outcome Treatment
1