FDA Adverse Event
Malfunction
Summary report: N
QUANTA SYSTEM S.P.A.
MDR report key: 3232308
·
Received December 21, 2012
Report
- Report Number
- 3004378299-2012-00024
- Event Type
- Malfunction
- Date Received
- December 21, 2012
- Date of Event
- November 27, 2012
- Report Date
- December 20, 2012
- Manufacturer
- QUANTA SYSTEM, S.P.A.
- Product Code
- GEX
- PMA / PMN Number
- K090962
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THE TWO CHILLER UNITS HAD USED UP THE THERMAL EXCHANGE GAS. THESE CHILLERS WERE REPLACED AND THE LASER SYSTEM RETURNED TO WORK. WE ARE UNAWARE ABOUT DELAY IN TREATMENT OR PT INJURY.
Description of Event or Problem · 1
THE TWO CHILLER UNITS DID NOT MAINTAIN THE TEMPERATURE SET POINT. WE ARE UNAWARE ABOUT DELAY ON TREATMENT OR PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | QUANTA SYSTEM S.P.A. | SURGICAL LASER | GEX | QUANTA SYSTEM, S.P.A. | CYBER TM 120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |