FREEDOM EVO-2 200
Report
- Report Number
- 3003402518-2013-00004
- Event Type
- Other
- Date Received
- July 15, 2013
- Date of Event
- June 24, 2013
- Report Date
- July 15, 2013
- Manufacturer
- TECAN SCHWEIZ AG
- Product Code
- JQW
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
A TECAN FIELD SERVICE ENGINEER WENT TO THE LABORATORY TO EVALUATE THE INSTRUMENT. THE DOOR LOCKS WERE DISABLED ALLOWING FOR EMPLOYEES TO REACH INTO THE INSTRUMENT WHILE IN OPERATION. TECAN INSTALLS INSTRUMENTATION WITH DOOR LOCKS ENABLED. ADDITIONALLY, TECAN OPERATOR'S MANUALS AND WARNING LABELS ON INSTRUMENT CAUTION OPERATORS TO NOT REACH INTO A RUNNING INSTRUMENT AS INJURY IS POSSIBLE. AFTER INVESTIGATION, IT IS UNCLEAR WHEN OR WHO DISABLED THE DOOR LOCKS. THE TECAN SERVICE ENGINEER ENABLED THE DOOR LOCKS. THE TECAN SERVICE ENGINEER ENABLED DOOR LOCKS AND INSTRUMENT PASSED APPLICABLE DIAGNOSTIC TESTING. NO EVIDENCE OF MALFUNCTION WAS OBSERVED.
AN EVO INSTRUMENT OPERATOR REACHED WITH THEIR HAND INTO THE INSTRUMENT WHILE IN OPERATION AND THE PICK AND PLACE ARM (PNP GRIPPER) HIT THEIR HAND PINNING IT TO THE INSIDE OF THE SAFETY SHIELD RESULTING IN A BRUISE. ON THE DAY OF THE EVENT, THE OPERATOR'S HAND WAS SWOLLEN AND HURTING. THERE WERE NO CUTS OR BROKEN SKIN AND THE OPERATOR DID NOT SEEK FIRST AID OR MEDICAL TREATMENT. THE OPERATOR WAS A CONTRACT EMPLOYEE OF THE LABORATORY AND INFORMED LABORATORY STAFF 3 DAYS LATER OF THE EVENT. THE LABORATORY REPORTED TO TECAN THE EVENT ON (B)(6). AT THAT TIME, THE OPERATOR INDICATED THE SWELLING HAD EXTENDED TO THE ARM. LAB STAFF ADVISED OPERATOR TO CONSULT A PHYSICIAN. NO FURTHER INFORMATION HAS BECOME AVAILABLE ABOUT THE NATURE OR EXTENT OF THE INJURIES OR TREATMENT REQUIRED. TECAN MADE MULTIPLE ATTEMPTS TO OBTAIN THIS INFORMATION. THE OPERATOR'S INJURY WAS THE RESULT OF DISABLED DOOR LOCKS WHICH ALLOWED THE OPERATOR TO REACH INSIDE A RUNNING INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 327664 | FREEDOM EVO-2 200 | PIPETTING STATION FOR CLINICAL USE | JQW | TECAN SCHWEIZ AG | EVO-2 200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |