FDA Adverse Event Malfunction Summary report: N

QUANTA SYSTEM

MDR report key: 3232276 · Received December 21, 2012

Report

Report Number
3004378299-2012-00027
Event Type
Malfunction
Date Received
December 21, 2012
Date of Event
September 21, 2012
Report Date
December 19, 2012
Manufacturer
QUANTA SYSTEM, S.P.A.
Product Code
GEX
PMA / PMN Number
K102749
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BROKEN OUTPUT COUPLER MIRROR. PLACED DISSECCAND INSIDE RESONATORS TO REDUCE HUMIDITY.

Description of Event or Problem · 1

THE SURGICAL LASER. AT THE LASER START THE LOW POWER ERROR IS DISPLAYED. IN SERVICE MODE CURRENT IS OK BUT POWER IN 0.0129 W AND NO OUTPUT FROM LASER HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTA SYSTEM SURGICAL LASER GEX QUANTA SYSTEM, S.P.A. CYBER TM 150

Patients

Seq Age Sex Outcome Treatment
1