FDA Adverse Event Injury Summary report: N

GORE DRYSEAL SHEATH

MDR report key: 3232267 · Received July 9, 2013

Report

Report Number
3007284313-2013-00056
Event Type
Injury
Date Received
July 9, 2013
Date of Event
June 10, 2013
Report Date
June 10, 2013
Manufacturer
W. L. GORE ASSOCIATES INC.
Product Code
DYB
PMA / PMN Number
K093791
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RESULT - THE REVIEW OF THE MFG PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECS. CONCLUSION - ACCORDING TO THE GORE DRYSEAL SHEATH WITH HYDROPHILIC COATING INSTRUCTIONS FOR USE (IFU) STATES ADEQUATE VESSEL ACCESS IS REQUIRED TO INTRODUCE THE SHEATH INTO THE VASCULATURE. CAREFUL EVAL OF VESSEL SIZE, ANATOMY, TORTUOSITY, AND DISEASE STATE (INCLUDING CALCIFICATION, PLAQUE, AND THROMBUS) IS REQUIRED TO ENSURE SUCCESSFUL SHEATH INTRODUCTION AND SUBSEQUENT WITHDRAWAL. IF VESSEL IS NOT ADEQUATE FOR ACCESS, MAJOR BLEEDING, VESSEL DAMAGE, OR SERIOUS INJURY TO THE PT, INCLUDING DEATH, MAY RESULT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE PT UNDERWENT AN ENDOVASCULAR PROCEDURE WITH GORE EXCLUDER AAA ENDOPROSTHESES TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE PT HAD SLIGHT TORTUOSITY AND HEAVY CALCIFICATION IN THE ILIAC ARTERIES. LEFT EXTERNAL ILIAC ARTERY MEASUREMENTS AT THE SMALLEST AREA WAS APPROX 6.4 TO 6.5MM. AT THE BEGINNING OF THE PROCEDURE, THE PHYSICIAN WAS INSERTING THE GORE DRYSEAL SHEATH (DSL 1828/11159476) INTO THE LEFT EXTERNAL ILIAC ARTERY. THE PHYSICIAN FELT SOME RESISTANCE UPON ADVANCING THE SHEATH BUT CONTINUED. IT WAS DISCOVERED THAT THE ILIAC ARTERY RUPTURED. BLOOD LOSS WAS MINIMAL. TWO GORE VIABAHN ENDOPROSTHESES WERE USED TO REPAIR THE ILIAC ARTERY. THE PT TOLERATED THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
313596 GORE DRYSEAL SHEATH DYB/INTRODUCER, CATHETER DYB W. L. GORE ASSOCIATES INC. 11159476

Patients

Seq Age Sex Outcome Treatment
1 67 YR Required Intervention