FDA Adverse Event Injury Summary report: N

ENDO CLIP II MED/LRG 10MM PISTOL GRIP

MDR report key: 3232242 · Received July 8, 2013

Report

Report Number
1219930-2013-00516
Event Type
Injury
Date Received
July 8, 2013
Date of Event
June 13, 2013
Report Date
June 14, 2013
Manufacturer
COVIDIEN, FORMERLY UNITED STATES
Product Code
FZP
PMA / PMN Number
K954435
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TRACKING NUMBER (B)(4).

Description of Event or Problem · 1

PROCEDURE: LADG - LAP ASSISTED DISTAL GASTRECTOMY. ACCORDING TO THE REPORTER: WHEN THE EXCISION ON THE LEFT GASTRIC ARTERY WAS PERFORMED, BLEEDING OCCURRED FROM UNDER THE CLIP ON THE PT SIDE. THE SURGEON CONSIDERS THE CLIP EDGE DAMAGED THE VESSEL WALL, RESULTING IN BLEEDING. THE VESSEL WAS LIGATED WITH SUTURE TO STOP THE BLEEDING. AMOUNT OF BLEEDING WAS LESS THAN 200 CC. THERE WAS NO UNANTICIPATED TISSUE LOSS. NOTHING FELL INTO CAVITY. PT IS UNDER OBSERVATION. OPERATING ROOM TIME NOT EXTENDED BY MORE THAN 30 MINUTES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308724 ENDO CLIP II MED/LRG 10MM PISTOL GRIP DISPOSABLE CLIP APPLIER FZP COVIDIEN, FORMERLY UNITED STATES

Patients

Seq Age Sex Outcome Treatment
1 74 YR Other