SETROX S 45
Report
- Report Number
- 1028232-2013-01949
- Event Type
- Malfunction
- Date Received
- July 17, 2013
- Date of Event
- July 9, 2013
- Report Date
- July 9, 2013
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVN
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
UPON RECEIPT, THE LEAD UNDER COMPLAINT WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. ITS PERFORMANCE WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION DEMONSTRATED THAT THE LEAD'S DISTAL PART WAS PULLED OUT OF THE RING ELECTRODE. THEREFORE, THE ADHESIVE RESIDUALS ON THE JOINING SURFACE OF LEAD TIP AND THE LEAD BODY WERE ANALYSED. THE ANALYSIS REVEALED NO INDICATION OF A MATERIAL OR A MANUFACTURING PROBLEM. AS THE ROOT CAUSE OF THE OBSERVED DAMAGE TRACTION FORCES DURING THE SURGERY SHOULD BE TAKEN INTO CONSIDERATION. DURING FURTHER ANALYSIS, NO OTHER DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
REP CALLED TO STATE THE PHYSICIAN WAS GOING TO PERFORM A LEAD REPOSITION DUE TO HIGH THRESHOLDS AND DIMINISHED ATRIAL SENSING VALUES. WE WERE INFORMED THE LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2013.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331088 | SETROX S 45 | PACER LEAD | NVN | BIOTRONIK SE & CO. KG | 350973 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Hospitalization |