FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3232106 · Received July 17, 2013

Report

Report Number
3008382007-2013-20152
Event Type
Malfunction
Date Received
July 17, 2013
Report Date
June 25, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Removal / Correction Number
2013-IE-017
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013 THE LAY USER/PATIENT IN (B)(6) CONTACTED LIFESCAN TO REPORT THE ONE TOUCH VERIO IQ METER WAS GIVING INACCURATELY HIGH READINGS COMPARED TO ANOTHER METER. THE PATIENT OBTAINED THE BLOOD GLUCOSE READING OF 181 MG/DL ON THE REPORTED METER AND THE READING OF 130 MG/DL ON THE OTHER METER. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED, THIS COMPLAINT IS BEING REPORTED. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331469 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL

Patients

Seq Age Sex Outcome Treatment
1 21 YR