FDA Adverse Event
Injury
Summary report: N
FINELINE II
MDR report key: 3232049
·
Received July 17, 2013
Report
- Report Number
- 2124215-2013-11357
- Event Type
- Injury
- Date Received
- July 17, 2013
- Date of Event
- June 19, 2013
- Report Date
- June 19, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
THE LEAD WAS SURGICALLY ABANDONED AND WILL NOT BE RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS ATRIAL LEAD WAS EXHIBITING IMPEDANCE MEASUREMENTS LESS THAN 200 OHMS AND OVERSENSING WHICH RESULTED IN PACING INHIBITION AND ASYSTOLE LESS THAN TWO SECONDS. A REVISION PROCEDURE WAS PERFORMED AND INSULATION DAMAGE WAS IDENTIFIED. THE LEAD WAS REMOVED FROM SERVICE WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331040 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4480 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 79 YR | Hospitalization| L | 4076| 4457| 1298| 4480| K063 |