Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE LEAD WAS SURGICALLY ABANDONED AND WILL NOT BE RETURNED FOR TESTING. THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM THE REPORTED CLINICAL OBSERVATIONS. NO OTHER ADVERSE EVENTS HAVE BEEN REPORTED. THIS PRODUCT ISSUE WILL BE UPDATED IF ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS ATRIAL LEAD WAS EXHIBITING IMPEDANCE MEASUREMENTS LESS THAN 200 OHMS AND OVERSENSING WHICH RESULTED IN PACING INHIBITION AND ASYSTOLE LESS THAN TWO SECONDS. A REVISION PROCEDURE WAS PERFORMED AND INSULATION DAMAGE WAS IDENTIFIED. THE LEAD WAS REMOVED FROM SERVICE WITHOUT FURTHER INCIDENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331040 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4480

Patients

Seq Age Sex Outcome Treatment
1 79 YR Hospitalization| L 4076| 4457| 1298| 4480| K063