FDA Adverse Event Malfunction Summary report: N

MEDRAD VERIS MR VITAL SIGNS MONITOR

MDR report key: 3231872 · Received July 15, 2013

Report

Report Number
2520313-2013-00038
Event Type
Malfunction
Date Received
July 15, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
MEDRAD
Product Code
MHX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BAYER R & I SERVICE REPLACED THE POWER CHARGER AND POWER SUPPLY CABLE. BAYER R & I PRODUCT ANALYSIS RECEIVED AND EXAMINED THE DC POWER CABLE AND POWER SUPPLY. THE EXAMINATION OF THE LEMO CONNECTOR AND THE ADJOINING WIRES DETERMINED THAT ONE OF THE WIRES HAD CROSSED OVER AND SHORTED WITH ANOTHER CONDUCTOR TO THE POINT WHERE THE INSULATION OF THE TWO CONDUCTORS BECAME CHARRED. A SUBSEQUENT INVESTIGATION OF SIMILAR COMPLAINTS CONCLUDED THAT WHEN CERTAIN COMBINATIONS OF CONDUCTORS SHORT TOGETHER AT THE CABLE END FARTHEST FROM THE POWER SUPPLY, THIS CAN CREATE A CONDUCTION PATHWAY THAT THE POWER SUPPLY DOES NOT RECOGNIZE AS A SHORT CIRCUIT AND SHUT DOWN. CONSEQUENTLY, THE POWER SUPPLY CONTINUES TO ENERGIZE THE CABLE WHICH CAN LEAD TO LOCALIZED HEATING AND MELTING OF THIS POWER CABLE. ON APRIL 15, 2013 BAYER HEALTHCARE DISTRIBUTED A FIELD SAFETY NOTICE RECALLING ALL 25-FOOT DC POWER CABLES SHIPPED WITH MEDRAD VERIS MR VITAL SIGNS MONITORS, OR THOSE PROVIDED BY BAYER SERVICE. THESE POWER CABLES ARE BEING RECALLED DUE TO A LATENT DESIGN RELIABILITY ISSUE AND THE POTENTIAL FOR SHORTING WHICH CAN RESULT IN HEATING/MELTING OF THE CABLE JACKET.

Description of Event or Problem · 1

THE SITE REPORTED THE FOLLOWING: THE CABLE FROM THE CHARGER IS WARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
326727 MEDRAD VERIS MR VITAL SIGNS MONITOR MONITOR, PHYSIOLOGICAL PATIENT MHX MEDRAD 3011994W

Patients

Seq Age Sex Outcome Treatment
1