FDA Adverse Event Malfunction Summary report: N

RESOLUTION CLIP CLIPPING DEVICE

MDR report key: 3231859 · Received July 17, 2013

Report

Report Number
3005099803-2013-06338
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
June 18, 2013
Report Date
June 18, 2013
Manufacturer
BOSTON SCIENTIFIC - MARLBOROUGH
Product Code
KOG
PMA / PMN Number
K040148
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SG
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

INVESTIGATION FOUND THAT THE DEVICE WAS FULLY DEPLOYED AND THE CLIP ASSEMBLY WAS NOT RETURNED. THE PRODUCT ANALYSIS ALSO INDICATES THAT THERE WAS A KINK IN THE CONTROL WIRE. BASED ON THE CONDITION OF THE RETURNED DEVICE, THE REPORTED FAILURE COULD NOT BE CONFIRMED. HOWEVER, DUE TO ANATOMICAL/PROCEDURAL FACTORS ENCOUNTERED DURING THE PROCEDURE, PERFORMANCE MAY HAVE BEEN LIMITED. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT.

Additional Manufacturer Narrative · 1

ALTHOUGH THE PATIENT AGE AND DATE OF BIRTH WERE NOT REPORTED, IT WAS REPORTED THE PATIENT WAS OVER 18 YEARS OLD. (B)(4) FOR THE REPORTED EVENT OF CLIP WOULD NOT RELEASE FROM CATHETER. ALTHOUGH THE SUSPECT DEVICE HAS BEEN RECEIVED, THE EVALUATION HAS NOT BEEN COMPLETED. THEREFORE, THE CAUSE OF THE REPORTED MALFUNCTION HAS NOT BEEN DETERMINED. UPON COMPLETION OF THE FAILURE ANALYSIS OF THE COMPLAINT DEVICE, IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MDR WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(6) 2013 THAT A RESOLUTION CLIP DEVICE WAS USED DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(6) 2013. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE RESOLUTION CLIP WAS DEPLOYED ONTO THE LESION; HOWEVER, WHEN THE OPERATOR REMOVED THE DEVICE, THE CLIP WAS STILL ATTACHED TO THE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER RESOLUTION CLIP DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION ON (B)(4) 2013 THAT A RESOLUTION CLIP DEVICE WAS USED DURING A GASTROSCOPY PROCEDURE PERFORMED ON (B)(4) 2013. ACCORDING TO THE COMPLAINANT, DURING THE PROCEDURE, THE RESOLUTION CLIP WAS DEPLOYED ONTO THE LESION; HOWEVER, WHEN THE OPERATOR REMOVED THE DEVICE, THE CLIP WAS STILL ATTACHED TO THE CATHETER. THE PROCEDURE WAS COMPLETED WITH ANOTHER RESOLUTION CLIP DEVICE. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. THE CONDITION OF THE PATIENT AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330528 RESOLUTION CLIP CLIPPING DEVICE ENDOSCOPE AND/OR ACCESSORIES KOG BOSTON SCIENTIFIC - MARLBOROUGH M00522611 ML000378C3

Patients

Seq Age Sex Outcome Treatment
1