OUTLOOK ES
Report
- Report Number
- 1641965-2013-00041
- Event Type
- Malfunction
- Date Received
- July 15, 2013
- Date of Event
- June 22, 2013
- Report Date
- June 25, 2013
- Manufacturer
- B. BRAUN MEDICAL, INC.
- Product Code
- FRN
- PMA / PMN Number
- K011975
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
AFTER THE ANALYSIS OF THE PUMP THE REPORTED COMPLAINT OF BATTERY EXPLOSION WAS NOT CONFIRMED; HOWEVER, THE OVERHEATING OF THE PUMP WAS DUPLICATED AND CONFIRMED. THE PUMP WAS OVERCHARGING CAUSING THE MAIN BATTERY TO OVERHEAT. THE BATTERY WAS CONTINUING TO CHARGE AT THE INITIAL CHARGE LEVEL OF 15 VOLTS INSTEAD OF REVERTING TO THE 14 VOLT MAINTENANCE CHARGE. BATTERIES WERE CHARGED FOR A DURATION OF 8 HOURS EACH THAT RESULTED IN THE OVERCHARGING OF THE BATTERY WHICH CAUSED THE OVERHEATING. THE PUMP WAS RAN AT 600 ML/HR FOR 2 HOURS DURING THE OVERCHARGING/OVERHEATING OF THE PUMP WITH NO VOLUMETRIC ERRORS OR ALARMS DISPLAYED. THE ROOT CAUSE WAS DETERMINED TO BE THE CHARGING IC, PART NUMBER (B)(4), ON THE MAIN BOARD OF THE UNIT. THE MAIN BOARD, (B)(4), AND MAIN BATTERY, (B)(4), WERE REPLACED TO CORRECT THE ISSUE. PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE SERVICE REPORTS ON THIS PUMP WAS PULLED DURING THE INVESTIGATION. THE PUMP HAD BEEN SENT IN FOR SERVICE ON 3 DIFFERENT OCCASIONS FOR LOSS OF THE DRUG LIBRARY. ON THE LAST DATE OF SERVICE, THE PUMPS MAIN BOARD AND COSMETIC BEZEL WAS REPLACED IN WHICH RESOLVED THE DRUG LIBRARY ISSUE.
THE CUSTOMER STATED THEY HAVE SENT THIS PARTICULAR PUMP IN ON NUMEROUS OCCASIONS FOR NOT UPLOADING CONFIGURATIONS. UPON THE RETURN OF THE PUMP THE CUSTOMER WAS PERFORMING PREVENTATIVE MAINTENANCE AND RUNNING MANY TESTS. AFTER RUNNING THE TESTS, THEY BEGAN VOLUMETRIC TESTING AND THEN THE PUMP BEGAN SOUNDING MANY ALARMS. SHORTLY THEREAFTER, THE PUMP BEGAN SMOKING AND WAS HOT TO THE TOUCH. THEY UNPLUGGED THE PUMP AND CHECKED THE BATTERY. THE BATTERY WAS REPLACED AND WAS DATED 2012. A FEW DAYS LATER THEY PLUGGED THE PUMP BACK IN AND ALLOWED IT TO RUN. A FEW HOURS LATER, THEY CHECKED THE PUMP AND IT AGAIN WAS SMOKING. THERE WERE NO INJURIES FROM THE EVENT AND NO PT INVOLVEMENT AS EVERYTHING TOOK PLACE IN THE BIOMED LAB.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 327777 | OUTLOOK ES | INFUSION PUMP | FRN | B. BRAUN MEDICAL, INC. | 621-400ES |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |