FDA Adverse Event Malfunction Summary report: N

PLV-100

MDR report key: 3231803 · Received July 15, 2013

Report

Report Number
2518422-2013-01398
Event Type
Malfunction
Date Received
July 15, 2013
Date of Event
May 18, 2011
Report Date
May 18, 2011
Manufacturer
RESPIRONICS INC.
Product Code
CBK
PMA / PMN Number
K832467
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, AN ISSUE RELATED TO THE LIMIT SWITCH AND PISTON CYLINDER WAS OBSERVED. THE DEVICE'S PISTON CYLINDER WAS REPLACED AND A LIMIT SWITCH WIRE WAS RESOLDERED TO ADDRESS THE ISSUE. THIS REPORT IS BEING SUBMITTED AS PART OF A PROGRAM TO RETROSPECTIVELY REVIEW POSSIBLE DEVICE MALFUNCTION THAT DID NOT INVOLVE ANY INJURY, TO DETERMINE WHETHER THEY ARE REPORTABLE UNDER 21 CFR 803 AND THE MANUFACTURER'S UPDATED REPORTING PROCEDURES. THIS PROGRAM IS BEING UNDERTAKEN TO ADDRESS FDA WARNING LETTER 12-PHI-01.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
326013 PLV-100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC. 35001

Patients

Seq Age Sex Outcome Treatment
1