FDA Adverse Event Injury Summary report: N

FABIUS GS

MDR report key: 3231801 · Received July 5, 2013

Report

Report Number
9611500-2013-00046
Event Type
Injury
Date Received
July 5, 2013
Date of Event
June 12, 2013
Report Date
July 1, 2013
Manufacturer
DRAEGER MEDICAL AG & CO. KG
Product Code
BSZ
PMA / PMN Number
K011404
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IF THE FRESH GAS HOSE IS NOT CONNECTED THE PRE USE CHECK WILL FAIL. THIS TEST IS REQUIRED TO BE PROCEEDED ACCORDING TO THE INSTRUCTIONS FOR USE ON A DAILY BASIS. IF THE FRESH GAS HOSE IS DISCONNECTED DURING A CASE THE VENTILATION WILL BE PERFORMED WITH ROOM AIR. DEPENDING ON THE OXYGEN UPTAKE FROM THE PATIENT AN INSPIRATORY OXYGEN LOW ALARM WILL BE POSTED IF THE SET ALARM LIMITS ARE EXCEEDED. THE LOWEST LIMIT THAT CAN BE SET IS 18%. THE LOGFILE OF THE FABIUS WAS SENT TO THE MFR FOR EVAL. THE LOGFILE INDICATED THAT AT 1:07 IN THE NIGHT A PRE USE CHECK WAS PERFORMED AND FAILED. LATER THAT DAY AT THE TIME OF THE REPORTED OCCURRENCE- THE FABIUS LOG FILE INDICATED A LOW OXYGEN CONCENTRATION (BELOW 10%). THIS ITEM IS RECORDED TWICE IN THE LOG WITHIN 2 MINUTES INTERVALS AND WILL LEAD TO A HIGH PRIORITY ALARM. AFTER FURTHER 8 MINUTES THE MESSAGE O2 LOW INOP IS ENTERED IN THE LOG. THERE EXISTS NO HINT TO A TECHNICAL DEVICE ERROR THAT ASSOCIATES TO THE REPORTED EVENT. PICTURES FROM THE ALARM LOGS WERE TAKEN AND ALSO SENT TO THE MFR. THE PICTURES WITH THE ENTRIES OF THE ALARM LOG WERE ANALYZED. THE ALARM LOG BOOK WAS CLEARED BEFORE THE INVESTIGATION COULD HAVE BEEN STARTED BY DRAGGER.

Description of Event or Problem · 1

IT WAS REPORTED THE USER FORGOT TO CONNECT THE FRESH GAS HOSE TO THE VENTILATION UNIT BECAUSE THE PREVIOUS CASE IT WAS OPEN SYSTEM CASE, THEN USER START NEXT OPERATION WITHOUT CONNECTION OF THE FRESH GAS HOSE TO THE VENTILATION UNIT COSY, WHICH LED TO A BRAIN DEATH OF THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308656 FABIUS GS ANAESTHESIA WORKSTATION BSZ DRAEGER MEDICAL AG & CO. KG NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK Life Threatening