FDA Adverse Event Malfunction Summary report: N

PANORAMA CENTRAL STATION WITH TELEPACK

MDR report key: 3231725 · Received July 12, 2013

Report

Report Number
2221819-2013-01434
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
January 22, 2013
Report Date
July 10, 2013
Manufacturer
MINDRAY DS USA, INC
Product Code
MHX
PMA / PMN Number
031760
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

COMPANY REP EVALUATED THE SYSTEM. CORRECTIONS INCLUDED REPLACEMENT OF THE SYSTEM'S WIRELESS TRANSCEIVER (TIM).

Description of Event or Problem · 1

CUSTOMER REPORTED THE PANORAMA CENTRAL STATION HAD LOST COMMUNICATION, WHICH MAY HAVE AFFECTED TELEMETRY MONITORING. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322520 PANORAMA CENTRAL STATION WITH TELEPACK CENTRAL STATION MONITOR MHX MINDRAY DS USA, INC

Patients

Seq Age Sex Outcome Treatment
1