LIGACLIP** ENDOSCOPIC CLIP APPLIER
Report
- Report Number
- 1527736-2013-00036
- Event Type
- Malfunction
- Date Received
- July 17, 2013
- Date of Event
- April 12, 2013
- Report Date
- June 10, 2013
- Manufacturer
- ETHICON ENDO SURGERY, INC (CINCINNATI)
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE ANALYSIS RESULTS OF THE DEVICE FOUND THAT IT WAS RECEIVED WITH THE JAWS MISALIGNED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON FIRING OF THE DEVICE, THE CLIPS WERE FED AS INTENDED; HOWEVER, DUE TO THE MISALIGNED CONDITION OF THE JAWS SCISSOR CLIPS WERE FORMED. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE WOULD NOT FIRE A CLIP. NO PATIENT IMPACT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331650 | LIGACLIP** ENDOSCOPIC CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO SURGERY, INC (CINCINNATI) | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |