CHARGING SYS
Report
- Report Number
- 1627487-2013-03975
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 19, 2013
- Report Date
- June 19, 2013
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS INCLUDED IN A FIELD CORRECTION. CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVALUATION RESULT: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPERATURE TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPGS AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 2 OF 2. REFERENCE MFR REPORT # 1627487-2013-03974. IT WAS REPORTED THE PT EXPERIENCED SCS IPG RECHARGE ISSUES AS WELL AS HER SCS IPG OVERHEATING. SUBSEQUENTLY, SURGICAL INTERVENTION WILL BE TAKEN AT A LATER DATE TO ADDRESS THE ISSUES. ADDITIONALLY, AT THIS TIME, IT IS UNK WHETHER OR NOT THE PT'S CHARGING SYS IS RELATED TO THE OVERHEATING ISSUE; THEREFORE, THE CHARGING SYS IS BEING REPORTED AS DEVICE 2. ON 08/01/2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 321629 | CHARGING SYS | SCS CHARGING SYS | LGW | ST JUDE MEDICAL - NEUROMODULATION | 3721 | 3415617 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | SCS ANCHOR: MODEL 1192 (2)| IMPLANT DATE:| SCS LEAD: MODEL 3286 (2)| IMPLANT DATE: |