Narratives
UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION WHICH OCCURRED MOST LIKELY DURING SURGERY. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD EXHIBITED LOSS OF CAPTURE AND LOW AMPLITUDE DUE TO LEAD DISLODGEMENT. THE LEAD WAS EXPLANTED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED.