EON MINI
Report
- Report Number
- 1627487-2013-03967
- Event Type
- Injury
- Date Received
- July 12, 2013
- Date of Event
- June 19, 2013
- Report Date
- June 19, 2013
- Manufacturer
- ST JUDE MEDICAL - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS IPG SERIAL NUMBER WAS INCLUDED IN A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT # 1627487-2013-03968. IT WAS REPORTED THE PT IS EXPERIENCING HIS SCS IPG HEATING WHILE CHARGING. SUBSEQUENTLY, THE PT REC'D A LOW ENERGY REPLACEMENT CHARGING SYS TO ADDRESS THE ISSUE. IT WAS ALSO REPORTED THE PT'S SCS IPG IS TURNING OFF AND ON BY ITSELF AND IS SOMETIMES NON-FUNCTIONAL. ON (B)(6) 2012, ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322107 | EON MINI | SCS IPG | LGW | ST JUDE MEDICAL - NEUROMODULATION | 3788 | 3558585 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 49 YR | Other | SCS LEAD: MODEL 3228| IMPLANT DATE:| IMPLANT DATE:| SCS ANCHOR: MODEL 1194 (2) |