FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 3231546 · Received July 17, 2013

Report

Report Number
3004209178-2013-11892
Event Type
Injury
Date Received
July 17, 2013
Date of Event
June 18, 2013
Report Date
June 24, 2013
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID 8709, LOT # J0039974R, IMPLANTED: (B)(6) 2000, PRODUCT TYPE CATHETER. (B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE PUMP REVEALED MOTOR GEAR TRAIN ANOMALY; CORROSION AND-OR WEAR AND-OR LUBRICATION. THE STALL WAS DUE TO SHAFT-BEARING.

Additional Manufacturer Narrative · 1

IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

THE PUMP PRINTOUT ALSO REVEALED THAT THERE WAS MORE THAN ONE STALL WITH RECOVERY AND A TUBE SET OCCURRED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A MOTOR STALLED OCCURRED. THE PUMP WAS INTERROGATED AND LOGS WERE READ VERIFYING THE PUMP STALLED ON (B)(6) 2013. THE PATIENT WAS ADMITTED TO THE HOSPITAL WITH POTENTIAL DRUG WITHDRAWAL AND THE PUMP WAS TO BE REPLACED. THE PATIENT EXPERIENCED FLU-LIKE SYMPTOMS INCLUDING THE CHILLS AND FEVER AND HAD ITCHING. THIS DEVICE SYSTEM DELIVERED INFUMORPH. ON (B)(6) 2013 ((B)(4)): IT WAS LATER REPORTED THAT THE PUMP WAS REPLACED (B)(6) 2013) AND THERE WAS NO PATIENT INJURY. THE PATIENT RECOVERED WITHOUT SEQUELA. ON (B)(6) 2013 ((B)(4)): NO NEW INFORMATION. ON (B)(6) 2013 (B)(4): NO NEW INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
331567 SYNCHROMED II PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MDT PUERTO RICO OPERATIONS CO 863740

Patients

Seq Age Sex Outcome Treatment
1 00050 YR Required Intervention