SYNCHROMED II
Report
- Report Number
- 3004209178-2013-11892
- Event Type
- Injury
- Date Received
- July 17, 2013
- Date of Event
- June 18, 2013
- Report Date
- June 24, 2013
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO
- Product Code
- LKK
- PMA / PMN Number
- P860004
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
CONCOMITANT PRODUCTS: PRODUCT ID 8709, LOT # J0039974R, IMPLANTED: (B)(6) 2000, PRODUCT TYPE CATHETER. (B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.
ANALYSIS OF THE PUMP REVEALED MOTOR GEAR TRAIN ANOMALY; CORROSION AND-OR WEAR AND-OR LUBRICATION. THE STALL WAS DUE TO SHAFT-BEARING.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
THE PUMP PRINTOUT ALSO REVEALED THAT THERE WAS MORE THAN ONE STALL WITH RECOVERY AND A TUBE SET OCCURRED.
IT WAS REPORTED THAT A MOTOR STALLED OCCURRED. THE PUMP WAS INTERROGATED AND LOGS WERE READ VERIFYING THE PUMP STALLED ON (B)(6) 2013. THE PATIENT WAS ADMITTED TO THE HOSPITAL WITH POTENTIAL DRUG WITHDRAWAL AND THE PUMP WAS TO BE REPLACED. THE PATIENT EXPERIENCED FLU-LIKE SYMPTOMS INCLUDING THE CHILLS AND FEVER AND HAD ITCHING. THIS DEVICE SYSTEM DELIVERED INFUMORPH. ON (B)(6) 2013 ((B)(4)): IT WAS LATER REPORTED THAT THE PUMP WAS REPLACED (B)(6) 2013) AND THERE WAS NO PATIENT INJURY. THE PATIENT RECOVERED WITHOUT SEQUELA. ON (B)(6) 2013 ((B)(4)): NO NEW INFORMATION. ON (B)(6) 2013 (B)(4): NO NEW INFORMATION.