FDA Adverse Event Injury Summary report: N

LAMITRODE S8

MDR report key: 3231533 · Received July 12, 2013

Report

Report Number
1627487-2013-10232
Event Type
Injury
Date Received
July 12, 2013
Date of Event
June 20, 2013
Report Date
June 21, 2013
Manufacturer
ST JUDE MEDICAL - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

METHOD - THE DEVICE HISTORY AND STERILIZATION RECORDS WERE REVIEWED. RESULTS - REVIEW OF THE DHR FOUND A NONCONFORMANCE RELATED TO THE PRODUCT LOT. HOWEVER, THE INDIVIDUAL AFFECTED DEVICE WAS REWORKED AND ALL DEVICES WITHIN THE LOT MET ACCEPTANCE CRITERIA. THEREFORE, THE DHR ANOMALY IS NOT RELATED TO THE ALLEGED DEVICE COMPLAINT. CONCLUSION - THE CAUSE OF THE REPORTED COMPLAINT COULD NOT BE DETERMINED FROM THE REVIEW OF THE DHR AND STERILIZATION RECORDS. SJM HAS LIMITED INFO RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.

Description of Event or Problem · 1

DEVICE 2 OF 2. REF MFR REPORT: 1627487-2013-10231. THE PT (B)(6) RECEIVED A TRIAL SCS SYSTEM WHICH INCLUDED A LEAD AND LEAD EXTENSION. IT WAS REPORTED WHEN THE PT RETURNED FOR THE PERMANENT IMPLANT, AN INFECTION WAS IDENTIFIED AT THE EXIT SITE OF THE LEAD EXTENSION. THE PHYSICIAN EXPLANTED THE EXTENSION AND LEFT THE LEAD IN-SITU. WOUND CULTURES WERE TAKEN AND THE PT WAS PLACED ON IV ANTIBIOTIC THERAPY. F/U INFO REVEALED CULTURE RESULTS RETURNED (B)(6) AND THE LEAD WAS EXPLANTED (EXPLANT DATE UNK). PLEASE NOTE: ADD'L DEVICE INFO FOR DEVICE 2 HAS BEEN REQUESTED; HOWEVER, NO FURTHER INFO WAS AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
324847 LAMITRODE S8 SCS LEAD LGW ST JUDE MEDICAL - NEUROMODULATION 3286 3951794

Patients

Seq Age Sex Outcome Treatment
1 UNK Required Intervention