DEXTRUS 4136
Report
- Report Number
- 1028232-2013-01906
- Event Type
- Injury
- Date Received
- July 17, 2013
- Date of Event
- May 16, 2013
- Report Date
- July 2, 2013
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- NVN
- PMA / PMN Number
- P950037
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE CONDUCTOR COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD HAD DISLODGED. THIS LEAD WAS REPOSITIONED BUT WAS REPORTEDLY DISLODGED AGAIN SO THE PHYSICIAN DECIDED TO REPLACE THIS LEAD. THIS LEAD WAS EXPLANTED AND OUT OF SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 330688 | DEXTRUS 4136 | PACER LEAD | NVN | BIOTRONIK SE & CO. KG | 358754 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |