FDA Adverse Event Malfunction Summary report: N

DEXTRUS 4137

MDR report key: 3231497 · Received July 17, 2013

Report

Report Number
1028232-2013-01915
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
May 6, 2013
Report Date
July 2, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE CONDUCTOR COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. IN THE COURSE OF THE FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS RIGHT VENTRICULAR LEAD EXPERIENCED A SYNCOPAL EPISODE WHILE WALKING UP A FLIGHT OF STAIRS. REVIEW OF STORED DEVICE MEMORY REVEALED AN EPISODE OF NOISE AND OVERSENSING OR A POTENTIAL EPISODE OF VENTRICULAR TACHYCARDIA. IT WAS UNABLE TO BE DETERMINED IF THE EPISODE WAS A TRUE VT, HOWEVER, THE PHYSICIAN FELT THAT IT WAS NOISE ON THE LEAD. AN INVASIVE PROCEDURE WAS PERFORMED. THE LEAD WAS SURGICALLY ABANDONED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330687 DEXTRUS 4137 PACER LEAD NVN BIOTRONIK SE & CO. KG 358764

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization