PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3231447 · Received July 17, 2013

Report

Report Number
2032227-2013-02922
Event Type
Injury
Date Received
July 17, 2013
Date of Event
June 14, 2013
Report Date
June 22, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURN FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP PASSED FUNCTIONAL TESTS INCLUDING DISPLACEMENT, PRIME, BASIC OCCLUSION, OCCLUSION AND NO DELIVERY TESTS. ALL BUTTONS FUNCTION PROPERLY. HOWEVER, MOISTURE DAMAGE FOUND ON KEYPAD TRACES.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE AND DKA. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF HOSPITALIZATION WAS OVER 600 MG/DL. CUSTOMER STATED THAT THE INFUSION SET WAS PULLED OUT OF THE SITE AND SHE DID NOT NOTICE UNTIL 16 HOURS LATER. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330542 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAL

Patients

Seq Age Sex Outcome Treatment
1 68 YR Hospitalization