FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3231227
·
Received July 17, 2013
Report
- Report Number
- 2032227-2013-02967
- Event Type
- Malfunction
- Date Received
- July 17, 2013
- Date of Event
- June 25, 2013
- Report Date
- June 25, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A LOOSE DRIVE SUPPORT DISK.
Description of Event or Problem · 1
THE CUSTOMER REPORTED AN ALARM DURING THE MANUAL PRIME. NO DAMAGE TO THE DRIVE SUPPORT CAP NOTED. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.