FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3231225 · Received July 17, 2013

Report

Report Number
2032227-2013-02965
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
June 25, 2013
Report Date
June 25, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP WAS UNABLE TO PRIME DURING THE PRIME TEST, AND ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST DUE TO A PROTRUDED/LOOSE DRIVE SUPPORT DISK. THE MOTOR PASSED THE MOTOR TEST. THE INSULIN PUMP HAD A MISSING END CAP STICKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP GAVE AN ALARM DURING THE PRIMING PROCESS. INSPECTED THE DRIVE SUPPORT CAP, AND IT WAS NOT DAMAGED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332199 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-722LCAL

Patients

Seq Age Sex Outcome Treatment
1