FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3231222 · Received July 17, 2013

Report

Report Number
2032227-2013-02961
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
March 23, 2013
Report Date
March 23, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSULIN PUMP FAILED THE DISPLACEMENT FOLLOWED BY A MOTOR ERROR ALARM DURING THE REWIND DUE TO AN OUT OF PHASE MOTOR ENCODER SIGNAL. THE INSULIN PUMP INVOLVED IN THIS EVENT IS THE PARADIGM REAL-TIME VEO INSULIN INFUSION PUMP, WHICH IS NOT MARKETED IN THE UNITED STATES. HOWEVER, THE DEVICE IS SIMILAR TO THE PARADIGM REAL-TIME INSULIN INFUSION PUMP, WHICH IS MARKETED IN THE UNITED STATES.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE REWIND AND MANUAL PRIME. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS NOT DAMAGED. UNABLE TO CONDUCT THE DISPLACEMENT TEST DUE TO THE MOTOR ERROR ALARMS. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
332198 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-754LCAH