FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT? PLUS

MDR report key: 3231166 · Received July 17, 2013

Report

Report Number
2134265-2013-04746
Event Type
Malfunction
Date Received
July 17, 2013
Date of Event
April 15, 2013
Report Date
June 20, 2013
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P110010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS A COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MANUFACTURER - EXAMINATION OF THE RETURNED DEVICE NOTED THAT THE HYOPTUBE WAS BROKEN AT APPROXIMATELY 49MM DISTAL TO THE STRAIN RELIEF. SEVERAL OTHER KINKS WERE NOTED ALONG THE LENGTH OF THE HYPOTUBE. A VISUAL AND MICROSCOPIC EXAMINATION OF THE STENT NOTED THAT TWO OF THE STENT STRUTS HAD LIFTED AT APPROXIMATELY 23MM PROXIMAL TO ITS DISTAL END. EXAMINATION OF THE FOLDED BALLOON MATERIAL AND THE DISTAL TIP SECTION FOUND NO ISSUES WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED (B)(4) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION, A SHAFT BEND OCCURRED. THE 88% STENOSED TARGET LESION WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE LESION WAS DILATED WITH A 1.5X15MM MAVERICK BALLOON. THE 2.5X38MM PROMUS ELEMENT STENT WAS ADVANCED HOWEVER WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER 2.5X38MM PROMUS ELEMENT STENT. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT STATUS IS STABLE, HOWEVER RETURNED DEVICE ANALYSIS REVEALED STENT DAMAGE AND A SHAFT FRACTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
330981 PROMUS ELEMENT? PLUS STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493918438250 15797488

Patients

Seq Age Sex Outcome Treatment
1 58 YR