FDA Adverse Event Malfunction Summary report: N

TERUMO TCM II COOLING AND HEATING SYSTEM

MDR report key: 3231118 · Received July 12, 2013

Report

Report Number
1828100-2013-00653
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
June 20, 2013
Report Date
June 20, 2013
Manufacturer
TERUMO CARDIOVASCULAR SYSTEMS CORP.
Product Code
DWC
PMA / PMN Number
K883603
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS COMPLAINT WAS CONFIRMED BY THE FIELD SERVICE REPRESENTATIVE (FSR). THE FSR FOUND A LOOSE HOSE CLAMP IN THE CARDIOPLEGIA PLUMBING CIRCUIT. WITH THE HOSE CLAMP TIGHTENED AND PREVENTATIVE MAINTENANCE COMPLETE, THE UNIT OPERATED TO MANUFACTURER SPECIFICATIONS AND WAS RETURNED TO CLINICAL USE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE ON THIS COMPLAINT THAT WOULD ALTER THE FACTS AND/OR CONCLUSION, A SUPPLEMENTAL REPORT WILL BE FILED ACCORDINGLY.

Description of Event or Problem · 1

THE SERVICE REPAIR TECHNICIAN (SRT) REPORTED THAT DURING ROUTINE TESTING OF THE DEVICE AT THE SERVICE CENTER, THERE WAS A WATER LEAK IN THE CARDIOPLEGIA CIRCUIT ON THE COOLER HEATER UNIT. THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
322284 TERUMO TCM II COOLING AND HEATING SYSTEM DWC TERUMO CARDIOVASCULAR SYSTEMS CORP. 4415

Patients

Seq Age Sex Outcome Treatment
1