FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3231044 · Received July 12, 2013

Report

Report Number
1720753-2013-08086
Event Type
Malfunction
Date Received
July 12, 2013
Date of Event
July 1, 2013
Report Date
July 12, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND ADJUSTED THE 5 VOLT POWER SUPPLY. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A COMMUNICATION FAILURE ERROR MESSAGE. THIS ERROR MAY RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN SITUATION DEPENDING ON WHEN THE LOSS OF COMMUNICATION OCCURS. THIS RESULTED IN A LOSS OF IMAGING FUNCTIONALITY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323238 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1