OT ULTRALINK METER
Report
- Report Number
- 3008382007-2013-20045
- Event Type
- Malfunction
- Date Received
- July 17, 2013
- Report Date
- July 6, 2013
- Manufacturer
- LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
- Product Code
- NBW
- PMA / PMN Number
- K073231
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PATIENT
Narratives
LIFESCAN (LFS) HAS REQUESTED RETURN OF THE SUBJECT PRODUCT(S) FOR EVALUATION. IF THE PRODUCT(S) ARE RETURNED, LFS WILL EVALUATE IT/THEM AND INFORM FDA OF PRODUCT(S) THAT DO NOT PASS INSPECTION IN A SUPPLEMENTAL REPORT.
ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING INACCURATE ERRATIC RESULTS. THE REPORTER OBTAINED BLOOD GLUCOSE VALUES OF 26, 54, AND 85MG/DL ON THE SAME LIFESCAN METER WITHIN 20 MINUTES OF EACH OTHER. THIS COMPLAINT IS BEING REPORTED FOR THE FOLLOWING REASON: BASED ON STATISTICAL METHODOLOGY, THE VARIATION BETWEEN THE RESULTS EXCEEDS THE MAXIMUM ACCEPTABLE VALUE OF 20% DIFFERENCE BETWEEN READINGS COMPARED TO ANOTHER METER WITHIN 20 MINUTES OF EACH OTHER. THE REPORTED RESULTS DID NOT MEET LIFESCAN'S ACCEPTABLE PRECISION CRITERIA AND THE ALLEGED INACCURACY ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT. REPLACEMENT PRODUCTS WERE SENT TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 331215 | OT ULTRALINK METER | GLUCOSE MONITORING SYS/KIT | NBW | LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL | 3343882 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |