PK DISSECTING FORCEPS INSTRUMENT
MDR report key: 3230669
·
Received July 16, 2013
Report
- Report Number
- 2955842-2013-02581
- Event Type
- Malfunction
- Date Received
- July 16, 2013
- Date of Event
- June 14, 2013
- Report Date
- June 17, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K061260
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING CONFIRMED THE REPORTED COMPLAINT. A FRAYED PITCH CABLE WAS FOUND AT THE INSTRUMENT WRIST. THE FRAYED SEGMENTS WERE IN VARIOUS LENGTHS. NO DAMAGE OR WEAR MARKS WERE FOUND AT THE CLEVIS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, THE CABLES WERE FRAYED ON A PK DISSECTING FORCEPS INSTRUMENT. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.