FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3230553 · Received July 16, 2013

Report

Report Number
3004209178-2013-95773
Event Type
Injury
Date Received
July 16, 2013
Date of Event
June 20, 2013
Report Date
June 24, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS. THE DEVICE HAD CRACKED CASE NEAR THE DISPLAY WINDOW CORNERS, CRACKED RESERVOIR TUBE LIP, AND CRACKED BATTERY TUBE THREADS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING IN THE EMERGENCY ROOM DUE TO DIABETES KETOACIDOSIS AND HIGH BLOOD GLUCOSE OF 820MG/DL. THE CUSTOMER STATED THAT HIS GLUCOSE LEVEL WAS HIGH ALL DAY AND HAD CHANGED THE INFUSION SET, BUT IT DID NOT DROP. THE CUSTOMER ALSO MENTIONED HAVING ISSUES WITH LOW BATTERY LATELY. THE CALLER STATED THAT THE INSULIN PUMP HAS CRACKS ON THE BATTERY COMPARTMENT. TROUBLESHOOTING WAS DECLINED AS CUSTOMER REQUESTED A REPLACEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
328217 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAL

Patients

Seq Age Sex Outcome Treatment
1 27 YR Hospitalization