PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3230546 · Received July 16, 2013

Report

Report Number
3004209178-2013-95758
Event Type
Injury
Date Received
July 16, 2013
Date of Event
June 20, 2013
Report Date
June 24, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
DC
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE ALARMS. UNABLE TO CONFIRM MOTOR ERROR ALARMS. THE MOTOR PASSED THE MOTOR TEST.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE OF ALMOST 800MG/DL. THE CALLER ALSO MENTIONED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS FLUSH. THE CALLER STATED THAT HE WOULD NOT BE RELEASED FROM THE HOSPITAL UNTIL HE RECEIVES A REPLACEMENT. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329148 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-722NAB

Patients

Seq Age Sex Outcome Treatment
1 28 YR Hospitalization