FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 3230546
·
Received July 16, 2013
Report
- Report Number
- 3004209178-2013-95758
- Event Type
- Injury
- Date Received
- July 16, 2013
- Date of Event
- June 20, 2013
- Report Date
- June 24, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DC
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST AS A RESULT OF A PROTRUDED/LOOSE DRIVE SUPPORT DISK. FURTHER TESTING COULD NOT BE PERFORMED DUE TO THE ALARMS. UNABLE TO CONFIRM MOTOR ERROR ALARMS. THE MOTOR PASSED THE MOTOR TEST.
Description of Event or Problem · 1
THE CUSTOMER REPORTED BEING HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE OF ALMOST 800MG/DL. THE CALLER ALSO MENTIONED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. TROUBLESHOOTING WAS PERFORMED, AND THE DRIVE SUPPORT CAP WAS FLUSH. THE CALLER STATED THAT HE WOULD NOT BE RELEASED FROM THE HOSPITAL UNTIL HE RECEIVES A REPLACEMENT. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329148 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-722NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR | Hospitalization |