FDA Adverse Event
Malfunction
Summary report: N
PARADIGM INSULIN INFUSION PUMP
MDR report key: 3230533
·
Received July 16, 2013
Report
- Report Number
- 3004209178-2013-95738
- Event Type
- Malfunction
- Date Received
- July 16, 2013
- Date of Event
- July 3, 2013
- Report Date
- July 12, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- LZG
- PMA / PMN Number
- K031390
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE, CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Additional Manufacturer Narrative · 1
THE INSULIN PUMP RECEIVED WITH LOOSE DRIVE SUPPORT DISK, SCRATCHED LCD WINDOW, CRACKED BATTERY TUBE THREADS, CRACKED RESERVOIR TUBE LIP, CRACKED BELT CLIP SLOT AND SCRATCHED RESERVOIR TUBE WINDOW.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE DRIVE SUPPORT CAP WAS PROTRUDED. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329405 | PARADIGM INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-712WWL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |