LARGE CLIP APPLIER INSTRUMENT
Report
- Report Number
- 2955842-2013-02573
- Event Type
- Malfunction
- Date Received
- July 16, 2013
- Date of Event
- June 11, 2013
- Report Date
- July 23, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT AND/OR ACCESSORY HAVE NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION), OR IF ADDITIONAL INFORMATION IS RECEIVED. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED COMPLAINT. UPON VISUAL INSPECTION, ENGINEERING FOUND NO BROKEN COMPONENTS AT THE DISTAL END. THE GRIPS OPENED AND CLOSED WHEN THE INPUTS WERE MANUALLY ROTATED. ENGINEERING OBSERVED THAT THE INSTRUMENT LUER PLATE WAS DISLODGED FROM THE CHASSIS AND PUSHED INTO THE BACKEND. A SECTION OF THE CHASSIS INNER WALL FEATURE THAT HELD THE LUER PLATE WAS BROKEN OFF. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO IMPROPER CLEANING. ENGINEERING FOUND THE CLAMPING PULLEY WAS CRACKED ON THE SCREW HEAD SIDE. THE SCREW HAD LOOSENED DUE TO THE CRACKING. ALL THE CLAMPING PULLEYS EXHIBITED WHITE RESIDUE AROUND THE CABLE GROOVES. THE EVIDENCE WAS INCONCLUSIVE, BUT THE DAMAGE WAS LIKELY DUE TO IMPROPER CLEANING. NO OTHER DAMAGE WAS FOUND. THE CLEANING AND STERILIZATION USER MANUAL SPECIFICALLY STATES: O WHEN SCRUBBING THE TIP OF CAUTERY INSTRUMENTS, TAKE CARE NOT TO DAMAGE THE INSULATION. O DO NOT EXPOSE INSTRUMENTS TO HYDROGEN PEROXIDE (H2O2), BLEACH, OR ALKALINE-BASED CLEANING AGENTS, AS THIS MAY RESULT IN INSTRUMENT DAMAGE. O PROLONGED EXPOSURE TO EITHER ULTRASONIC CLEANING OR CLEANING AGENTS MAY RESULT IN INSTRUMENT DAMAGE. RECURRENCE OF THE ALLEGED FAILURE MODE IS NOT EXPECTED TO CAUSE OR CONTRIBUTE TO A DEATH, SERIOUS INJURY OR SERIOUS DETERIORATION IN THE STATE OF HEALTH OF A PATIENT. BASED ON THIS ADDITIONAL FAILURE ANALYSIS INVESTIGATION INFORMATION, THIS MDR REPORT IS BEING RETRACTED.
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE PERMANENT CAUTERY SPATULA INSTRUMENT WAS NOTED TO HAVE A BROKEN TIP. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 329434 | LARGE CLIP APPLIER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420230-03 | 2009072 361 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DAVINCI SI SYSTEM, ESU, INSTRUMENTS & ACCESSORIES |