FDA Adverse Event Malfunction Summary report: N

5MM, 33CM L-TIP ELECTROSURGICAL PROBE (3BX)

MDR report key: 3230495 · Received July 16, 2013

Report

Report Number
0002936485-2013-00301
Event Type
Malfunction
Date Received
July 16, 2013
Date of Event
April 9, 2013
Report Date
June 27, 2013
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
GCJ
PMA / PMN Number
K934094
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS NOT RETURNED FOR INVESTIGATION. THEREFORE, THE REPORTED FAILURE MODE COULD NOT BE CONFIRMED. THE POSSIBLE ROOT CAUSES FOR THE REPORTED FAILURE CAN BE DUE TO: USE ERROR, IMPROPER STERILIZATION AND/OR NORMAL WEAR AND TEAR. HOWEVER, THIS CANNOT BE CONFIRMED SINCE THE UNIT WAS NOT RETURNED. IN THE EVENT THAT THE UNIT IS RETURNED, A FULL EVALUATION WILL BE CONDUCTED AND A FOLLOW UP REPORT WILL BE ISSUED. IN SUM, THE PRODUCT WAS NOT RETURNED FOR INVESTIGATION AND THE REPORTED FAILURE MODE COULD NOT BE CONFIRMED. THE FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON THOUGHT HE SAW A SPARK JUST PROXIMAL TO THE HOOK TIP. THE SPARK OCCURRED WHILE HOOK WAS BEING USED TO FREE THE GALLBLADDER FROM THE LIVER AND THUS WAS IN CLOSE PROXIMITY TO THE LIVER. THE DEVICE WAS REMOVED IMMEDIATELY AND THE SCRUB NURSE CONFIRMED THAT THERE WAS A DEFECT IN THE INSULATION. SURGEON INSPECTED THE LIVER BED, DUODENUM, AND COLON AND OBSERVED NO INJURY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SURGEON THOUGHT HE SAW A SPARK JUST PROXIMAL TO THE HOOK TIP. THE SPARK OCCURED WHILE HOOK WAS BEING USED TO FREE THE GALLBLADDER FROM THE LIVER AND THUS WAS IN CLOSE PROXIMITY TO THE LIVER. THE DEVICE WAS REMOVED IMMEDIATELY AND THE SCRUB NURSE CONFIRMED THAT THERE WAS A DEFECT IN THE INSULATION. SURGEON INSPECTED THE LIVER BED, DOUDENUM, AND COLON AND OBSERVED NO INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329137 5MM, 33CM L-TIP ELECTROSURGICAL PROBE (3BX) LAPAROSCOPE, GENERAL & PLASTIC SURGERY GCJ STRYKER ENDOSCOPY-SAN JOSE UNK

Patients

Seq Age Sex Outcome Treatment
1