FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3230167 · Received July 16, 2013

Report

Report Number
2531779-2013-11005
Event Type
Malfunction
Date Received
July 16, 2013
Report Date
July 10, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 10/01/2013 - DEVICE EVALUATION:THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS 09/10/2013 WITH THE FOLLOWING FINDINGS: DURING TESTING, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. A TEST SCREEN WAS INSERTED AND WAS FOUND TO FUNCTION PROPERLY. UNRELATED TO THE COMPLAINT, EVALUATION REVEALED A CRACK ALONG THE BATTERY COMPARTMENT EXTENDING FROM THE THREADS TO THE CASE SEAL.

Additional Manufacturer Narrative · 1

ANIMAS HAS REQUESTED THE SUBJECT PRODUCTS TO BE RETURNED FOR EVALUATION HOWEVER THE PRODUCTS HAVE NOT YET BEEN RETURNED OR EVALUATED. IF ANIMAS RECEIVES THE PRODUCT, ANIMAS WILL INFORM THE FDA OF ANY PRODUCTS THAT FAIL EVALUATION IN A SUPPLEMENTAL REPORT.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THE REPORTER ALLEGED THAT THE ISSUE OCCURRED GRADUALLY AND THE CONTRAST WAS ALREADY SET AT 10. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329967 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 76 YR