FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3230057 · Received July 16, 2013

Report

Report Number
2531779-2013-10990
Event Type
Malfunction
Date Received
July 16, 2013
Report Date
July 1, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1: DATE OF SUBMISSION 10/23/2013-DEVICE EVALUATION:THE INSULIN PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/14/2013 WITH THE FOLLOWING FINDINGS:ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP SHOWED SOME COSMETIC DEFECTS AND NO DAMAGE TO THE KEYPAD WAS OBSERVED. ON INVESTIGATION, ALL THE BUTTONS WERE RESPONDING PROPERLY. REMOVAL OF THE KEYPAD COVER DID NOT REVEAL ANY CONTAMINATION OR DAMAGE TO THE BUTTON CONTACTS. PUMP COVER WAS REMOVED AND EXAMINATION OF THE KEYPAD FLEX DID NOT SHOW ANY ANOMALY.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER ALLEGED THAT THE KEYPAD BUTTONS WERE DIFFICULT TO PRESS. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
329689 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 12 YR