FDA Adverse Event Malfunction Summary report: N

GOBED II MED/SURG BED

MDR report key: 3230028 · Received July 16, 2013

Report

Report Number
0001831750-2013-06388
Event Type
Malfunction
Date Received
July 16, 2013
Date of Event
June 20, 2013
Report Date
June 21, 2013
Manufacturer
STRYKER MEDICAL-KALAMAZOO
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCLUSION: DEVICE TO BE REPLACED.

Additional Manufacturer Narrative · 1

FOLLOW-UP SUBMITTED AS FURTHER INVESTIGATION DETERMINED THE FOWLER WAS STUCK IN A LOW, FLAT POSITION. THIS WILL RESULT IN CAREGIVER ANNOYANCE; HOWEVER, IT IS NOT LIKELY TO HARM THE PATIENT AS THIS POSITION IS DESIRED FOR CPR ADMINISTRATION, IF REQUIRED. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED. THIS ISSUE IS NOT LIKELY TO CAUSE OR CONTRIBUTE TO SERIOUS INJURY OR DEATH IF IT WAS TO REOCCUR.

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOWLER WOULD NOT LAY FLAT DUE TO BENT FRAME AND BRACES. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Description of Event or Problem · 1

IT WAS REPORTED VIA REPAIR WORK ORDER THAT THE FOWLER WOULD NOT LAY FLAT DUE TO BENT FRAME AND BRACES. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.