FDA Adverse Event Other Summary report: N

DELTA TT CUP, DIA.56 MM

MDR report key: 3229810 · Received July 4, 2013

Report

Report Number
3008021110-2013-00010
Event Type
Other
Date Received
July 4, 2013
Date of Event
May 27, 2013
Report Date
July 4, 2013
Manufacturer
LIMACORPORATE S.P.A.
Product Code
LPH
PMA / PMN Number
K112898
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS IS THE 1ST SIMILAR CASE REPORTED TO LIMACORPORATE. WE CHECKED THE DHR OF THE DELTA TT CUP INVOLVED, WITHOUT FINDING ANY ANOMALY WHICH COULD HAVE CAUSED THE PRESENCE OF A SHAVING ON THE CUP EXTERNAL SURFACE. AT THE MOMENT, WE RECEIVED A PHOTO OF THE CUP INVOLVED, WHICH SHOWS THE PRESENCE OF THE SHAVING ON THE POLAR PART OF THE CUP EXTERNAL SURFACE. BASED ON THIS PHOTO, IT APPEARS THAT THE SHAVING COULD COME FROM THE MACHINING PROCESS OF THE CUP. WE WILL ALSO RECEIVE THE CUP, THEREFORE, WE WILL SUBMIT A FOLLOW-UP REPORT AFTER ANALYSIS ON IT.

Description of Event or Problem · 1

THIS IS AN INTRA-OPERATIVE ISSUE, OCCURRED IN (B)(6). THE DELTA TT CUP WAS OPENED IN THEATRE, AND FOUND TO HAVE WHAT LOOKED LIKE A HAIR BUT ON CLOSER EXAMINATION IT APPEARS TO BE FINE SHAVINGS ON ITS EXTERNAL SURFACE. THERE WAS NO PATIENT INVOLVEMENT, AS THE SHAVING WAS EASILY IDENTIFIED, AND THE CUP HAS NOT BEEN USED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306671 DELTA TT CUP, DIA.56 MM LPH, MBL LPH LIMACORPORATE S.P.A. 5552.15.560 201304062

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Hospitalization