FDA Adverse Event Death Summary report: N

PRUITT-INAHARA CAROTID SHUNT

MDR report key: 322943 · Received March 28, 2001

Report

Report Number
MW1021472
Event Type
Death
Date Received
March 28, 2001
Date of Event
March 15, 2001
Manufacturer
IDEAS FOR MEDICINE
Product Code
DQR
Adverse Event
Yes
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

PHYSICIAN INSERTED THE INTERNAL CAROTID SHUNT INTO THE COMMON CAROTID ARTERY AND THE COMMON CAROTID SHUNT INTO THE INTERNAL CAROTID ARTERY. WHEN THE BALLOON WAS INFLATED IT RUPTURED THE INTERNAL CAROTID AT THE BASE OF THE SKULL. SUGGEST COLOR CODING THE ENDS OF THE SHUNTS AND PUTTING A PRESSURE RELEASE ON THE COMMON CAROTID SHUNT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
13749 PRUITT-INAHARA CAROTID SHUNT CAROTID SHUNT, OUTLYING, MODEL DQR IDEAS FOR MEDICINE 400-40-9F 1021999-11

Patients

Seq Age Sex Outcome Treatment
1 78 YR Death