FDA Adverse Event
Death
Summary report: N
PRUITT-INAHARA CAROTID SHUNT
MDR report key: 322943
·
Received March 28, 2001
Report
- Report Number
- MW1021472
- Event Type
- Death
- Date Received
- March 28, 2001
- Date of Event
- March 15, 2001
- Manufacturer
- IDEAS FOR MEDICINE
- Product Code
- DQR
- Adverse Event
- Yes
- Report Source
- Voluntary report
- Reporter Location
- MI, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
PHYSICIAN INSERTED THE INTERNAL CAROTID SHUNT INTO THE COMMON CAROTID ARTERY AND THE COMMON CAROTID SHUNT INTO THE INTERNAL CAROTID ARTERY. WHEN THE BALLOON WAS INFLATED IT RUPTURED THE INTERNAL CAROTID AT THE BASE OF THE SKULL. SUGGEST COLOR CODING THE ENDS OF THE SHUNTS AND PUTTING A PRESSURE RELEASE ON THE COMMON CAROTID SHUNT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 13749 | PRUITT-INAHARA CAROTID SHUNT | CAROTID SHUNT, OUTLYING, MODEL | DQR | IDEAS FOR MEDICINE | 400-40-9F | 1021999-11 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Death |