INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00529
- Event Type
- Malfunction
- Date Received
- July 14, 2013
- Report Date
- June 19, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K034026
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE COMPLAINT RT235 CIRCUIT IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.
(B)(4) METHOD: THE RETURNED COMPLAINT RT235 BREATHING CIRCUIT WAS VISUALLY INSPECTED AND PRESSURE TESTED FOR THE REPORTED LEAK. RESULTS: THE PRESSURE TEST REVEALED THAT THE RT235 CIRCUIT WAS WITHIN SPECIFICATION AND WAS NOT LEAKING. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR LOT 130305. CONCLUSION: WE WERE UNABLE TO DETERMINE WHAT CAUSED THE LEAK REPORTED BY THE CUSTOMER AS OUR TESTING CONFIRMED THAT THERE WAS NO DEFECT WITH THE SUBJECT BREATHING CIRCUIT. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT235 INFANT DUAL-HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARM." (B)(4).
A HOSPITAL IN (B)(6) REPORTED THAT AN RT235 INFANT BREATHING CIRCUIT HAD A LEAK. THIS WAS FOUND DURING THE VENTILATOR LEAK TEST, BEFORE PATIENT USE.
A HOSPITAL IN (B)(6) REPORTED THAT AN RT235 INFANT BREATHING CIRCUIT HAD A LEAK. THIS WAS FOUND DURING THE VENTILATOR LEAK TEST, BEFORE PATIENT USE.