FDA Adverse Event Injury Summary report: N

LOCKSCR Ø3.5 SELF-TAP L40 TAN

MDR report key: 3225560 · Received July 13, 2013

Report

Report Number
8030965-2013-04280
Event Type
Injury
Date Received
July 13, 2013
Date of Event
June 18, 2013
Report Date
June 19, 2013
Manufacturer
SYNTHES GMBH
Product Code
KTT
PMA / PMN Number
K000684
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND NO COMPLAINT RELATED ISSUES WERE FOUND. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.

Additional Manufacturer Narrative · 1

THIS DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. THE ADDITIONAL EVALUATION FOUND THAT THE MEASURABLE DIMENSIONS WERE IN COMPLIANCE WITH THE TECHNICAL DRAWINGS. THE EXAMINATION OF THE RAW-MATERIAL TESTING CERTIFICATES AND THE MANUFACTURING PAPERWORK SHOWED NO DEVIATIONS REGARDING MATERIAL ANALYSIS, STRENGTH AND STRUCTURAL STABILITY. BASED ON PROVIDED DETAILS THIS EVENT IS MOST LIKELY THE RESULT OF TOO MUCH MECHANICAL FORCE DURING REMOVAL. MICROSCOPIC VIEW OF THE BROKEN SURFACES DO NOT SHOW ANY ANOMALIES OF MATERIAL STRUCTURE. NO PRODUCT FAULT COULD BE DETECTED. NOTE: BLANK FIELDS ON THIS FORM INDICATE INFORMATION THAT IS UNKNOWN, UNAVAILABLE OR UNCHANGED. PLACEHOLDER.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN THE (B)(6) AS FOLLOWS: WHILE TRYING TO REMOVE THE PLATE FROM THE PATIENT WITH SIX MONTHS NON-UNION DISTAL MEDICAL FRACTURE, THE LOCKING SCREWS BROKE FROM THE NECK OF THE SCREWS. THERE WERE THREE BROKEN SCREWS. THE PLATE WAS NOT BROKEN. NO XRAY WAS TAKEN BEFORE THE PROCEDURE. THE PATIENT WAS COMPLAINING OF PAIN THAT IS WHY THE SURGEON DECIDED TO REMOVE THE PLATE AND LOOKED FOR ANOTHER SOLUTION. THE PATIENT DETAILS WERE NOT GIVEN BECAUSE OF THE RESTRICTION IN THE MILITARY HOSPITAL. IF FURTHER DETAILS ARE NEEDED, THE SURGEON CAN BE CONTACTED. INVOLVED ARTICLES: 1X DMT PLATE 439.901 LOT. 3801008, 2X SCREW 413.040 LOT. 7812930 BROKEN, 1X SCREW 406.045 LOT. 3802833. MATERIAL RECEIVED ON (B)(4) 2013. THIS IS REPORT 2 OF 5 FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
325127 LOCKSCR Ø3.5 SELF-TAP L40 TAN KTT SYNTHES GMBH 7812930

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention