FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3223933 · Received July 5, 2013

Report

Report Number
3008642652-2013-01793
Event Type
Malfunction
Date Received
July 5, 2013
Date of Event
June 14, 2013
Report Date
July 3, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLES) HAS BEEN CONFIRMED. UPON INVESTIGATION CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS SEVERED. THE ROOT CAUSE FOR THE SEVERED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE SEVERED CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A NURSE CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT A (B)(6) FEMALE PT'S ELECTRODE BELT HAD DAMAGED CABLES. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
308245 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR